Measurement radiation level at Atati, Butajera, Mercy and Woliso hospital was carried out using THERMO/RadEye PRD (personal radiation detector). The radiation level results obtained range from 3.074±0.144623mSv/y to 4.179±0.141461mSv/y with an average of 3.6205±0.13417mSv/y respectively. These results showed that the radiation levels in all of the internal (inside) radiation room were higher than the recommended values for patient that 1 mSv/y maximum permissible limit for the public set by International Commission on Radiological Protection (ICRP).